Instrument Acronym:
DCALS (Dual Channel Airborne Laser Spectrometer)
Instrument Measurements:
a) Trace Gas
Formaldehyde
b) Cloud Physics/Aerosol
c) Radiation/Optical/Electrical
d) Additional Capabilities
Resource Person/Developer:
| Name: | Alan Fried | Bryan P. Wert |
| Affiliation: | ACD | ACD |
| Phone: | (303) 497-1475 | (303) 497-1451 |
| Email: | fried@ucar.edu | bwert@ucar.edu |
Principle of Measurement:
Infrared Absorption
In situ/Remote/Flux Measurement:
in situ
Field projects in which the instrument has been used:
NARE 96, STERAO A
Aircraft Platform:
NOAA P3
Inlet Configuration/Location:
standard inlet, rear-facing, with heated winglet
Calibration/Cross Calibration:
Gas phase standards at about 5 ppbv level
every 20 min, at inlet entrance
Operator Requirements:
one
Data Collection:
self
Post-flight Data Validation, Processing, Hours per Flight Hour etc.
Data fitting and inspection of data quality
Physical Characteristics:
Dimensions: 40x30x36 in
Weight: 450-500 lbs
Power: 5 amp, 110 V, 60 Hz
Other:
Response Time:
20 sec to 3 min, depending upon desired signal to noice ratio
Detection Limits:
80 pptv to 40 pptv for the above times
Accuracy:
5% of ambient concentration + detection limit
Precision:
at 1000 times the detection limit precision is approximately 0.1 - 0.2%
Limitations (such as altitude, airspeed, etc.)
none known